Clean & Sterilize System
The cleaning and sterilization of the system is very simple. There are only a few parts to sterilize; handpiece and cable (detached from the SonicWelder Rx), Sonotrode tip – removable by wrench (included), Xcelsior hot waterbath cover and basin (optional), twist drills and sonotrode tips. All of the implants are already sterile, so there are very few items to prepare!
Sterilization method and parametersThe following parameters are recommended: For pre-vacuum cycles, a 4-minute exposure time at 270°F. Caution: Time required for sterilizer to reach temperature is not included in the times given. This is based on instructions in Steam Sterilization and Sterility Assurance in Health Care Facilities (ANSI/AAMI ST46-2002-5.8.1 and 5.8.2.)
Please reference the product insert for complete instructions on SonicWeld Rx and other devices and implants.
Safety note:
The responsibility for proper cleaning, disinfection and resterilization rests with the operator/user of the product. Be sure to observe your local regulations, including any process related restrictions. For sterilizationand subsequent transportation and storage, approved sterilization packaging (e.g. as specified by EN 868, ISO 11607) must be used as appropriate.